Catalog contour Ts By English
Your Meter Overview 2
Your Meter Features 4
Information For Safety 5
Getting Ready to Test 7
Getting the Blood Drop 9
Testing Your Blood 10
Cleaning Your Meter 13
Control Solution Testing 14
Setting the Time, Date and Sound 18
Transferring Results to a Computer 21
Viewing Stored Results 22
Error Codes and Symbols 24
Changing the Battery 28
Caring for Your Meter 29
Symptoms of High or Low Blood Glucose 30
System Specifications 31
Service Information 37
Warranty 38
Supplies 40
INTENDED USE: The CONTOURTMTS system (meter, test strips, controls) is intended for self-testing by people with diabetes and by health care professionals to monitor glucose concentrations in fresh capillary whole blood drawn from the fingertip. In clinical settings
the CONTOUR TS system may also be used with fresh venous and arterial whole blood. The test provides a quantitative measurement of glucose in blood from 10 mg/dL to 600 mg/dL
The CONTOUR TS blood glucose monitoring system is not intended for diagnosis of or screening for diabetes mellitus, and is not for use in neonatal care
Information For Safety
Read your CONTOURTMTS user guide, the lancing device insert and all instructional materials provided in your meter kit before testing. Please follow all instructions for use and care
exactly as described to help avoid inaccurate results.
Wash and dry your hands well before testing.
Used test strips and lancets are possible biohazards and must be
disposed of as medical waste or as advised by your health care professional.
The lancing device provided with your kit is intended for self
testing by a single patient. It must not be used by more than one person due to the risk of infection.
Warning Potential Biohazard
Health care professionals or persons using this system on multiple patients should follow the infection control procedure approved by their facility. All products or objects which come in contact with human blood, even after cleaning, should be handled as if capable of transmitting infectious diseases. The user should follow the recommendations for prevention of blood borne transmissible diseases in health care settings as recommended for potentially infectious human specimens.
For complete cleaning instructions, see page 13.
Your Meter Display
All display segments will appear for 10 seconds. If you need more than 10 seconds to check the display, press and hold again
While all characters are displayed, it is very important to verify that 8.8.8 is fully displayed. If there are missing segments, see Error Codes and Symbols (pages 24 – 27). This may affect the way you see your results.
Thank you for choosing the CONTOURTMTS blood glucose monitoring system. We are proud to be your partner in helping you manage your diabetes
Your Meter Features
Easy: The CONTOUR TS blood glucose monitoring system is about simplicity. You will notice its simple approach the first time you see it.
Automatic: Your CONTOUR TS meter is designed with No Coding technology, to automatically code itself every time you insert a test strip. It will even tell you when a test strip is under filled.
NO CODING
Programmed: Your CONTOUR TS meter has been preset at the factory with the standard date and time for your convenience. . Your meter has been preset and locked to display results
in mg/dL (milligrams of glucose per deciliter). In some countries, the meters are preset to display units in mmol/L
(millimoles of glucose per liter) Results in mg/dL will never have a decimal point
Results in mmol/L will always have a decimal point. Example
96 mg dl or 5.3 mmol/L
Check your display to ensure that results are being displayed in mg/dL. If they are not, contact Customer Service. See the back cover for contact information.
Information For Safety
Always keep the test strips in the original bottle Tightly close the bottle immediately after removing
a test strip. The bottle is designed to keep the test strips dry. Exposure to room humidity from leaving the bottle open or not storing them in the original bottle can damage your test strips. This could lead
to inaccurate test results.
Wash and dry your hands well before testing. • ® Test strips are for single use only. Do not
use a test strip that appears damaged or has been used
Your test strip is designed to easily “sip” the
blood into the sample tip. Do not drop blood directly on the flat surface of the test strip
Do not press the test strip against your finger when testing. This may block the sample tip
Let the meter and test strips adjust to the temperature where you are testing.
Whenever the meter is moved from one location to another, allow approximately 20 minutes for the meter to adjust to the temperature of a new location before performing a blood glucose test. The meter has been designed to give
accurate results at temperatures between 5°C and 45°C.
Check the expiry dates on your test strips and control solution. It is important to not use the test strips or control solution if the expiry date printed on the bottle label and carton has passed. For the control solution do not use if it has been six months since you first opened the bottle. It will help to write the six month discard date on the control solution label.
The following steps are shown in the proper order for performing a blood glucose test, pages 8 – 12.
Preparing the Test Strip
1. Wash and dry your hands well before testing.
2. Remove a CONTOURTMTS test strip from the bottle.
3. Hold the test strip with the grey end facing up.
4. Insert the grey end into the test strip port on the meter.
The meter will turn on. An image of a test strip with a flashing blood drop will appear letting you know the meter is ready to test Preparing the Lancing Device
The lancing device graphics are for illustration purposes only. Your lancing device may look different. Refer to your lancing device insert for detailed instructions on preparing the lancing device
The lancing device provided with your kit is intended for self testing by a single patient. It must not be used on more than one person due to the risk of infection.
Dispose of the used lancet as medical waste or as advised by your health care professional.
Do not reuse lancets. Use a new lancet each time you test
Information For Safety
Have all the materials you will need ready before you begin testing. This includes your CONTOURTMTS meter, the CONTOURTMTS test strips, the lancing device and lancets. You will also need CONTOURTMTS control solution to run a quality control check. CONTOUR TS control solutions are sold separately.
. The lancing device provided with your kit is intended for self-testing by a single patient. It must not be used on more than one person due to the risk of infection.
Use a new lancet each time you test because it is no longer sterile after use.
Do not perform a blood glucose test when the CONTOUR TS meter is connected to an external device,
e.g., computer.
Use only approved equipment from the manufacturer or certified body such as UL or TUV
Examine product for missing, damaged, or broken parts. If the test strip bottle is open inside a new box of strips, do not use those strips. Contact Customer Service for replacement parts. See the back cover forget contact information.
All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even after you have performed cleaning. See page 13. – Always wash your hands well with soap and water and dry them before and after testing, handling the meter, lancing device or test strips.
1 . Remove the end cap from the lancing device.
2 . Loosen the round protective cap on a lancet by rotating it 1/4 turn, but do not remove it.
3 . Insert the lancet firmly into the lancing device until it comes to a full stop
4 . Twist off the round protective lancet cap. Save it for disposing of the used lancet.
5 . Replace the end cap.
6 . Rotate the end cap dial to adjust the puncture depth. The amount of pressure applied to the puncture site also affects puncture depth
Getting the Blood Drop and Testing
1. Press the lancing device firmly against the puncture site and press the release button.
2 . Stroke your hand and finger towards the puncture site to form a drop of blood. Do not squeeze around the puncture site.
3 . Test immediately after a good blood drop has formed
4 . Immediately touch the tip of the test strip to the drop of blood. The blood is pulled into the test strip through the tip.
Hold the tip of the test strip in the blood drop until the meter beeps. Do not press the tip against the skin or place the blood on top of the test strip or you could get inaccurate results or errors.
5 . After a beep, you will see the meter count down 5 seconds and your blood test result is displayed and automatically stored in the meter memory. Please do not touch the test strip during the countdown as this may result in an error.
6 . Remove the test strip to turn your meter off. Dispose of the used test strip as medical waste or as advised by your health care professional.
Test Results Expected Values
Blood glucose values will vary depending on food intake, medication dosages, health, stress, or activity. Non-diabetic plasma glucose concentrations are normally maintained within a relatively narrow range, roughly 70 – 110 mg/dL in a fasting state. You should consult with your health care professional for expected values specific to your needs.
If your blood glucose reading is under 50mg/dL, or you see (less than 10 mg/dL) on the meter display, follow medical advice immediately. If your test result is above 250 mg/dL, or you see H 1 (greater than 600 mg/dL) on the meter display, call your health care professional as
soon as possible. . Always consult your health care professional
before changing your medication based on CONTOURTMTS test results.
Ejecting and Disposing of the Used Lancet
1. Do not use your fingers to remove the lancet from the lancing device. The lancing device provided with your kit has an automatic lancet ejection feature.
2. Refer to the separate lancing device insert, if provided with your kit, for instructions on automatic ejection of the lancet
Always dispose of the used test strip and lancet as medical waste or as advised by your health care professional.
Do not reuse lancets. Use a new lancet each time you test
Cleaning Your Meter
The exterior of the CONTOURTMTS meter can be cleaned using a moist (not wet) lint free tissue with a mild detergent or disinfectant solution, such as 1 part bleach mixed with 9 parts water. Wipe dry with lint free tissue after cleaning.
Do not allow cleaning solution to run in or around the button, battery cover or test strip port. Doing so may cause a malfunction.
Control Solution Testing
Use only CONTOURTMTS control solutions (Normal, Low or High) with your CONTOURTMTS blood glucose monitoring system.
Using anything other than CONTOUR TS control solution could present incorrect results
You should perform a control test:
When using your meter for the first time
When you open a new bottle or package of test strips
If you think your meter may not be working properly
If you have repeated unexpected blood glucose results
Normal, Low or High control solutions are available and sold separately. Always use CONTOUR TS control solutions. Other brands could present incorrect results. If you need help locating CONTOUR TS control solutions, contact Customer Service. See the back cover for contact information.
1. Wash and dry your hands well before testing.
2. Remove a test strip from the bottle and firmly snap the
lid closed
NOTE: Check the expiry date on the test strip bottle and the expiry and discard dates on control solution. Do not use expired materials.
3. Hold the test strip with the grey end facing up.
4. Insert the grey end into the test strip port on the meter.
The meter will turn on. An image of a test strip with a flashing blood drop will appear letting you know the meter is ready to test.
5 . Gently rock the control bottle before opening to ensure itis mixed well.
6 . Squeeze a small drop of control solution on a clean, nonabsorbent surface, such as a piece of wax paper. Do not apply control solution to your fingertip or to the test strip directly from the bottle.
7 . Immediately touch the tip of the test strip to the drop of control solution. The solution is pulled into the test strip through the tip.
8. Hold it in the drop until the meter beeps.
You will see the meter count down 5 seconds until the test is complete and your control test result is displayed.
The meter will automatically recognize and ✓ mark the control result for you.
NOTE: The does not indicate the control test result is within the specified range. It only marks it as a control solution test. It will not be included in your averages.
9. Compare your control test result to the ranges provided on the test strip bottle or carton.
If your control test result is out of range, do not use your meter for blood glucose testing until you resolve the issue. See “What You Should Do” on page 27.
To turn your meter off, simply remove the test strip. Dispose of the used test strip as a medical waste.
The result will be automatically ✓marked and stored in the meter memory. Control results will not be included in your 14 day blood glucose average.
Your CONTOURTMTS meter allows you to set the date, time, and sound as you like them.
You can view 12 or 24 hour time
month/day or day. month, and;
choose to have your meter beep or not
To get into set-up mode when the meter is off, press and hold M for 3 seconds A complete display will briefly appear, then numbers will flash in the bottom right corner of the display.
Setting the Time
Setting the Time The number in the hour position will be flashing. Press ♥ until you reach the correct hour. Press M to set
The number in the minute position will now flash. Press ♥ until you reach the correct minutes. Press M to set
Setting the Date
the year will be flashing press ♥ to select the current year Press M to set
The month will now flash. Press ♥ to select the month. Press M to set
The day will now flash. Press ♥ to select the day. Press M to set
The flashing bell symbol will now appear on the display. This option controls whether you want to hear the beep sound during testing. Press M to set
Setting the 12 or 24 Hour Clock
24H will now flash on the display
Press ♥ to choose the 12H or 24H setting This option lets you set the meter to display time as a 12 hour setting with AM and PM or as a 24 hour display
Press M to set
Setting the Month and Day Format
d.m will now flash on the display. Press ♥ to select the format you wish (m/d for month/day format or d.m for day. Month format). Press M to set
When you reach the end of the setup, your meter screen will display all of the current settings and turn off automatically.
Transferring Results to a Computer
You can transfer test results from the CONTOURTMTS meter to a computer, where they can be summarized in a report with graphs and tables. To make use of this feature, you need GLUCOFACTS DELUXE diabetes management software and a USB data cable. The cable is not commercially available but can be requested from Customer Service. See the back cover for contact information. For a free download of the software go to www.glucofacts.ascensia.com or contact Customer Service. See the back cover for contact information. May not be available in all countries.
IMPORTANT: The CONTOUR TS meter has not been tested by Ascensia Diabetes Care for use with any software other than diabetes management software from Ascensia Diabetes Care. Ascensia Diabetes Care is not responsible for any erroneous results from the use of other software.
Viewing Stored Test Results
With the meter off, press and release M to view stored test results. The 14 day average will appear in the center of the display. Any control results are not included in this 14 day average
Press M to review previous results. Results will be shown starting with the most recent. Each result will show the date and time the test was taken. You can press to scroll through the results.
When End appears on the display, you have viewed all of the results in the memory.
Press M to turn the meter off, or it will automatically turn off after 3 minutes
NOTES
Averages include the specified time period plus the
Current day. The meter will hold 250 results in the memory. When the
Maximum is reached, the oldest test result will be removed as a new test is completed and saved to the memory.
Press M to turn the meter off, or it will turn off automatically after three minutes.
Low Battery – when battery symbol remains on display screen during testing
Replace the battery as soon as battery symbol possible (see page 28)
Temperature out-of-range.
Move to an area that is within the operating range for the meter: 5°C-45°C. Let the meter and test strips adjust to this temperature for twenty minutes before testing.
The test strip and repeat filled enough to give an accurate result
You may have applied blood more than once on the same strip.
Blocked sample tip…
Blood drop too small.
· Remove the test using a new test strip.
· Do not attempt to apply blood more than once per strip
· Do not push strip into skin while filling strip
· See suggested drop size on page 9.
The meter is sensing Remove the test strip
Remove the test strip and retest a used test strip. using a new test strip. Wait until you see the flashing blood drop in the display before adding the blood sample.
Test strip not inserted correctly.
Remove the test strip and insert correctly it properly (see page 7). Use only CONTOURTMTS test strips
Potential software or hardware issue
Remove the test strip and retest using a new test strip. If you continue to have problems, contact Customer Service. See the back cover for contact information
Wrong test strip.
Remove the test strip and retest using only a CONTOUR TS Test Strip.
Strip disturbed during countdown.
Strip degraded due to long exposure out of bottle.
Used test strip inserted.
Remove the test strip and retest using a new test strip. Carefully follow testing instructions in your user guide..
if you continue to have problems, contact Customer Service. See the back cover for contact information.
Changing the Battery
The battery should be replaced when the meter continually displays or does not turn on. Your CONTOURTMTS meter uses one 3-volt lithium battery (DL2032 or CR2032)
1. Turn off your meter before changing your battery.
2. Press firmly on the battery cover and slide in the direction of the arrow.
NOTE: After replacing the battery you will need to reset the time (see page 18). The date and other formatting choices do not need to be reset. Test results stored in the memory are not erased when the battery is changed.
3 . Remove the old battery by placing your finger under the top of the battery and pulling upward. Grasp the battery and remove
4. Place new battery into the battery compartment with the “+” side up.
5 . Slide the battery cover back into place, lining up with the open slots, and close firmly.
CAUTION
Keep battery cover and battery out of reach of children. Accidental swallowing could cause suffocation. Lithium batteries are poisonous. If swallowed, immediately contact your doctor or poison control center. Discard batteries according to your local environmental regulations.
Changing the Battery
Caring for Your Meter
Store meter in the carrying case provided whenever
possible.
Wash and dry hands well before handling to keep
the meter and test strips free of water, oils and other contaminants. Handle the meter carefully to avoid damaging the
electronics or causing other malfunctions.
Avoid exposing meter and test strips to excessive humidity, heat, cold, dust, or dirt.
Clean your meter as recommended on page 13.
Symptoms of High or Low Blood Glucose
You can better understand your test results by being aware of the symptoms of high or low blood glucose. According to the American Diabetes Association, some of the most common symptoms are
Low blood glucose (Hypoglycaemia):
shakiness
sweating
fast heartbeat
blurred vision
confusion
passing out
seizure
irritability
extreme hunger
dizziness
High blood glucose (Hyperglycaemia):
frequent urination
excessive thirst
blurred vision
increased fatigue
hunger
Ketones (Ketoacidosis)
shortness of breath
nausea or vomiting
very dry mouth
Information For Safety
If you are experiencing any of these symptoms, test your blood glucose. If your test result is under 50 mg/dL or above 250 mg/dL, contact your health care professional immediately.
For additional information and a complete list of symptoms, contact your health care professional.
System Specifications
Test Sample: Venous, arterial or capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 uL
Measuring Range: 10-600 mg/dL
Results: 5-second countdown
Memory Feature: Stores most recent 250 test results
Battery Type: One 3-volt lithium battery, 225 mAh capacity
(DL2032 or CR2032)
Battery Life: Approximately 1000 tests (1 yr. average use)
Operating Temperature Range: r*
Humidity: 10% – 93% RH 5°C
Dimensions: 71 mm (H) x 60 mm (W) x 19 mm (T)
Weight: 56.7 grams
Sound: A beep sounds whenever a test strip is inserted into the
meter, when the test strip is filled with blood or control solution, or when a test result appears on the display screen. Two beeps will sound for an error. To turn off the sound, see Setting the Sound on page 19
Electromagnetic Compatibility (EMC): The CONTOURTMTS meter complies with the electromagnetic requirement specified in ISO 15197:2013. Electromagnetic emissions are low and unlikely to interfere with other nearby electronic equipment, nor are emissions from nearby electronic equipment likely to interfere with the CONTOUR TS meter. The CONTOUR TS meter meets the requirements of IEC 61000-4-2 for immunity to electrostatic discharge. Avoid use of electronic devices in very dry environments, especially if synthetic materials are present. The CONTOUR TS meter meets the requirements of IEC 61326-1 for radio frequency interference. To avoid radio frequency interference, do not use the CONTOURTMTS meter near electrical or electronic equipment that are sources of electromagnetic radiation, as these may interfere with the proper operation of the meter.
Comparison Options: The CONTOURTMTS system is designed for use with venous and capillary whole blood. Comparison to a laboratory method must be done simultaneously with aliquots of the same sample. Note: Glucose concentrations drop rapidly due to glycolysis (approximately 5% – 7% per hour).
The following symbols are used throughout the product labeling for the CONTOURTMTS blood glucose monitoring system (meter packaging and labeling, and test strip and control solution packaging and labeling).
Symbols Used
Use by date (last day of month)
Do not reuse
Batch code
Temperature limitations
In Vitro Diagnostic Medical Device
Manufacturer
Caution
Catalogue number
CONTROL number
Control Range Normal
Control Range Low
Control Range High
Control Discard Date
Sterilized using irradiation
Batteries must be disposed of in accordance with laws in your country. Contact your competent local authority for information on the relevant laws regarding disposal and recycling in your area. The meter should be treated as contaminated and disposed of according to local safety rules. It should not be disposed of with waste electronic equipment. Contact your health care professional or local waste disposal authority for medical waste disposal guidelines.
Accuracy
The CONTOURTMTS blood glucose monitoring system was tested in 100 capillary blood samples using 600 CONTOURT
MTS test strips. Two replicates were tested with each of 3 lots of CONTOUR TS test strips for a total of 600 readings. Results were compared to the YSI® glucose analyzer, which is traceable to the CDC hexokinase method. The tables below compare the performance of the 2 methods.
Table 1 – System accuracy results for glucose concentration < 100 mg/dL
Difference range in values between YSI laboratory reference method and CONTOUR TS meter |
Within + 5 mg/dL |
Within + 10 mg/dL |
Within + 15 mg/dL |
|
|
|
|
Table 2 – System accuracy results for glucose concentration – 100 mg/dL
Difference range in values between YSI laboratory reference method and CONTOUR TS meter |
Within + 5% |
Within + 10% |
Within + 15% |
Number (and percent) of samples within specified range |
297 of 420 (70.7%) |
399 of 420 (95.0%) |
| 416 of 420 (99.0%) |
Table 3 – System accuracy results for glucose concentrations between 36 and 443 mg/dL
Within 15 mg/dL or +15% |
586 of 600 (97.7%) |
Acceptance criteria in ISO 15197: 2013 are that 95% of the measured glucose values shall fall within either +15 mg/dL of the average measured values of the reference measurement procedure at glucose concentrations <100 mg/dL or within +15 % at glucose concentrations 2 100 mg/dL
User Accuracy A study evaluating glucose values from fingertip capillary blood samples obtained by 132 lay persons showed the following results: 100% within +15 mg/dL of the medical laboratory values at glucose concentrations below 100 mg/dL and 95.6% within +15% of the medical laboratory glucose concentrations at or above 100 mg/dL.
Precision
A repeatability study was conducted with the CONTOURTMTS blood glucose monitoring system using 5 venous whole blood specimens with glucose levels from 40 to 329 mg/dL. Multiple replicates (n=300) were tested using multiple CONTOUR TS blood glucose meters and 3 lots of CONTOUR TS blood glucose test strips. The following precision results were obtained.
Table 4 – System repeatability results for CONTOURTMTS meter using CONTOURTMTS test strips
Mean, mg/dL |
Pooled Standard Deviation mg/dL |
95% Confidence Interval of Standard Deviation, mg/dL |
Coefficient of Variation, % |
39.6 |
1.4 |
1.31 – 1.55 |
3.6 |
76.1 |
1.8 |
1.70 – 2.01 |
2.4 |
131.8 |
2.8 |
2.58 – 3.06 |
2.1 |
199.4 |
3.4 |
3.10 – 3.67 |
1.7 |
326.9 |
4.8 |
4.47 – 5.30 |
1.5 |
Intermediate precision (which includes variability across multiple days) was evaluated using 3 control solutions. With each of 3 lots of CONTOUR TS test strips, each control was tested once on each of 10 instruments on 10 separate days for a total of 300 readings.
Table 5 – System intermediate precision results for CONTOUR TS meter using CONTOUR TS test strips
Control Level |
Mean, mg/dl |
Standard Deviation, mg/dL |
95% Confidence Interval of Standard Deviation, mg/dL |
Coefficient of Variation, % |
Low |
39.9 |
0.06 |
0.57 – 0.67 |
1.5 |
Normal |
122.4 |
1.6 |
1.49 – 1.76 |
1.3 |
high |
358.1 |
4.5 |
4.15 – 4.91 |
1.3 |
Service Information If you have a problem and none of the problem solving steps in this user guide help, contact Customer Service. See the back cover for contact information. We have trained specialists to help you
Important
Speak to a Customer Service Representative before returning your meter for any reason. He/she will give you the information needed to get your problem handled correctly and efficiently. Have your CONTOURTMTS blood glucose meter and CONTOURTMTS test strips available when you phone. It would also be helpful to have a bottle of CONTOURTMTS control solution nearby.
Check List
This check list may be helpful when speaking with Customer Service
1. Locate the serial number (A) and model number (B) on the back of the meter.
3 . Check the battery icon on the display. (See page 28,
Changing the Battery)
Manufacturer’s Warranty Diabetes Care warrants to the original purchaser that this instrument will be free from defects in materials and workmanship for 5 years from the date of original purchase (except as noted below). During the stated 5-year period,
Ascensia Diabetes Care shall, at no charge, replace a unit found to be defective with an equivalent or current version of the owner’s model. Limitations of Warranty:This warranty is subject to the following exceptions and limitations:
1. A 90-day warranty only will be extended for consumable
parts and/or accessories.
2. This warranty is limited to replacement due to defects
in parts or workmanship. Ascensia Diabetes Care shall not be required to replace any units that malfunction or are damaged due to abuse, accidents, alteration, misuse, neglect, maintenance by someone other than Ascensia Diabetes Care, or failure to operate the instrument in accordance with instructions. Further, Ascensia Diabetes Care assumes no liability for malfunction of or damage to Ascensia Diabetes Care instruments caused by the use of test strips or control solution other than the appropriate products recommended by Ascensia Diabetes Care (i.e., CONTOURTMTS test strips and CONTOURTMTS control
Solutions).
3. Ascensia Diabetes Care reserves the right to make
Changes in the design of this instrument without obligation to incorporate such changes into previously manufactured instruments.
4. Ascensia Diabetes Care has no knowledge of the
performance of the CONTOURTMTS blood glucose meter when used with any test strips other than CONTOUR TS test strips, and therefore does not warrant
CONTOURTMTS meter CONTOURTMTS test strips
the performance of the CONTOURTMTS meter when used with any test strips other than CONTOURTMTS test strips or when the CONTOUR TS test strip is altered or modified
in any manner.
5. Ascensia Diabetes Care makes no warranty regarding
the performance of the CONTOUR TS meter or test results when used with any control solution other than
CONTOURTMTS control solution.
6. Ascensia Diabetes Care makes no warranty regarding
the performance of the CONTOUR TS meter or test results when used with any software other than the CONTOUR DIABETES APP (where supported) or the GLUCOSE FACTS DELUXE diabetes management software
(where supported) from ASCENSIA DIABETES CARE. ASCENSIA DIABETES CARE MAKES NO OTHER EXPRESS WARRANTY FOR THIS PRODUCT. THE OPTION OF REPLACEMENT, DESCRIBED ABOVE, IS THE ONLY OBLIGATION OF ASCENSIA DIABETES CARE UNDER THIS WARRANTY. IN NO EVENT SHALL ASCENSIA DIABETES CARE BE LIABLE FOR INDIRECT, SPECIAL OR CONSEQUENTIAL DAMAGES, EVEN IF ASCENSIA DIABETES CARE HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. Some States do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. This warranty gives you specific legal rights and you may also have other rights which vary from State to State.
For warranty service: Purchaser must contact Ascensia Diabetes Care Customer Service for assistance and/or instructions for obtaining service of this instrument. See the back cover for contact information.
You can contact Customer Service for supplies. See the back cover for contact information.
REPLACEMENT PARTS
Item
One 3-volt lithium battery DL2032 or CR2032
CONTOURTMTS user guide
CONTOURTMTS test strips
CONTOURTMTS control solution, Normal
CONTOURTMTS control solution, Low
CONTOURTMTS control solution, High
REFERENCES
1. Sewell DL. Protection of Laboratory Workers From Occupationally Acquired Infections; Approved Guideline, 3rd Edition. Clinical and Laboratory Standards Institute. CLSI document M29-A3; ISBN 156238-567-4. March 2005.
2. Cryer PE, Davis SN. Hypoglycemia. In: Kasper D, et al, editors. Harrison’s Principles of Internal Medicine. 19th edition. New York, NY: McGraw Hill; 2015
http://accessmedicine.mhmedical.
com/content.aspx?bookid=1130§ionid=79753191
3. American Diabetes Association. http://www.diabetes.org
4. Burtis CA, Ashwood ER, editors. Tietz Fundamentals of Clinical
Chemistry. 5th edition. Philadelphia, PA: WB Saunders Co; 2001;444.